Tingkat Selektivitas, Faktor Kapasitas, Jumlah Pelat Teoritik dan Waktu Retensi Analisis PAH dengan GC-FID Menggunakan Kolom RTX-5-MS dan Kolom CP-Sil 8 CB

Muhammad Syahrir(1*),

(1) Jurusan Kimia FMIPA UNM Makassar
(*) Corresponding Author




DOI: https://doi.org/10.35580/chemica.v13i2.628

Abstract


Kromatografi Gas tipe QP 2010 mampu mendeteksi Poliaromatik Hidrokarbon (PAH) skala nanogram permikroliter. Jenis kolom yang dipakai memerlukan suatu ketepatan dan ketersesuaian yang bergantung pada PAH. Beberapa jenis PAH yang dipakai dalam membandingkan jenis kolom dalam penelitian ini adalah naftalena, asenaftene,fenantrena, florantena,pirena, Benzoantrasena dan perilena. Membandingkan dua jenis kolom Kolom Rtx-5-MS dan Kolom CP-Sil 8 CB. Tingkat selektivitas, faktor kapasitas, jumlah pelat teoritis serta waktu retensi menjadi kriteria dalam diamati. Hasil penelitian menunjukkan bahwa kolom Rtx-5-MS pada kondisi kromatografi gas memberikan tingkat selektifitas, faktor kapasitas dan jumlah pelat teoritis memenuhi syarat pemisahan yang baik dan waktu retensi antara 4,60 menit – 30,00 menit, sedangkan kolom CP-Sil 8 CB      (25 m x 0,23 mm) pada waktu retensi antara 4,60 menit – 25,00 menit, sedikit lebih cepat dalam terdeteksi PAH dengan tingkat selektivitas yang lebih tinggi. Kolom Rtx-5-MS memenuhi syarat pemisahan PAH dengan baik tetapi memerlukan waktu kurang lebih 5 menit lebih lama munculnya peak Perylene yang terpisah, menyulitkan menganalisis jenis PAH tersebut dibandingkan dengan pemakaian kolom CP-Sil-8 CB.

Kata Kunci: selektivitas, faktor kapasitas,pelat teoritik, GC-FID, PAH

ABSTRACT

Gas chromatography QP 2010 type is able to detect the polyaromatic hydrocarbons (PAH) on nanogram per microlitre scale. The types of column that was used require a precision and appropriate that depend on PAH. Several types of PAH used in comparing column types on this study is naphtalene, acenaphtene, phenantrene, fluoranthene, pyrene, benzoantracene and perylene. Comparing two types of colums, RTX-5-MS column and CP-Sil 8 CB colum. Degree of selectivity, capacity factors, number of theoretical plates and retention times become criterion were observed. The results showed that RTX-5-MS column on gas chromatography provide degree of selectivity, capacity factors and theoretical plates are qualified for a good separation and retention time of 4.60 min – 30.00 min, while CP-Sil 8 CB (25 m x 0,23 mm) at the retention times of 4.60 min – 25.00 min, slightly faster in PAH detection with higher selectivity. RTX-5-MS column is qualified in PAH separation but it takes about 5 minutes longer to Perylene peak separate, it makes difficult to analyze PAH types compared CP-Sil-CB colum.

Keywords: selectivity, capacity factors, theoretical plates, GC-FID, PAH


Full Text:

PDF

Article Metrics

Abstract view : 1697 times | PDF view : 432 times

Refbacks

  • There are currently no refbacks.


ISSN: 2722-8649

Diterbitkan oleh Jurusan Kimia FMIPA Universitas Negeri Makassar (UNM) dua kali dalam setahun yaitu bulan Juni dan Desember.

Alamat Redaksi :Jurusan Kimia, Fakultas MIPA UNM, JL. Dg. Tata Parangtambung, Makassar 90224 Indonesia, Telp. 0411-840295; Fax: 0411840295;E-mail : chemica@unm.ac.id